8-4 VS 9/VS 8 Vital Signs Monitor Service Manual
NOTE
To ensure correct upgrade, set the IP address of the PC to 77.77.1.xx before
running the upgrade program. There are no special requirements for the IP
addresses of the gateway and DNS. For example, the IP address is
77.77.1.13, and the subnet mask is 255.255.255.0.
8.2.1.3
Upgrading Software Tool
System software upgrade settings:
1. Download the large software package of the monitoring system (the storage
location is the model package path), run the installed system (network) upgrade tool
software, click
Select New Model Package
, select the
LEGO.Tool
model package, click
Open
, and then click
OK
, as shown in the following figures:
2. In the
Product Type Selection
dialog box, select
LEGO
as the machine type.
The following figure shows the PC screen:
8.2.2
Upgrade Through CMS
This monitor can be upgraded through the BeneVision central monitoring system (CMS).
Note: The CMS version must be up-to-date if you want to upgrade the monitor software
through the CMS. The operation screens vary with the CMS version. For details, see the
CMS operator’s manual.
Содержание VS8
Страница 1: ...VS9 VS8 Vital Signs Monitor Service Manual ...
Страница 2: ......
Страница 8: ...VI VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 36: ...3 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 62: ...4 26 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 116: ...6 46 VS 9 VS 8 Vital Signs Monitor Service Manual 6 4 7 8 E 4 4 Removing the Touchscreen Assembly Same as E 4 3 ...
Страница 140: ...8 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 155: ......
Страница 156: ...PN 046 020105 00 2 0 ...