VS 9/VS 8 Vital Signs Monitor Service Manual
6-15
6.3.5.4
(C-4) Removing the Panel Assembly
1. Unplug the NIBP air tube at the position shown in the figure on the panel.
2. Unplug the CO
2
air tube and cable indicated by the arrows on the panel, and unplug
the SpO
2
cable and CO
2
cable on the main board; (subject to the actual
configuration of the monitor). Replace the VS9 panel assembly or SpO
2
cable
separately;
Rotate the SpO
2
cable clockwise to remove it. The new SpO
2
cable must be connected at
the correct position.