A-8 VS 9/VS 8 Vital Signs Monitor Service Manual
monitor.
WPA2-PSK is highly
recommended.
WPA2-Enterprise may
increase the offline
probability while
roaming, so it is not
recommended.
department of the
hospital.
4
Private
virtual local
area
network
(VLAN)
The monitor needs to use
the private VLAN.
The VLAN minimizes
broadcast or multicast
data that may affect
stability of the monitor.
Check if this
requirement is satisfied
together with the IT
department of the
hospital.
Important settings
1
DHCP
The DHCP server needs to
reserve enough IP
addresses to ensure that
the monitor can obtain
the IP addresses.
Check if this
requirement is satisfied
together with the IT
department of the
hospital.
2
IGMP
snooping
If the monitor adopts
multicast, the IGMP
snooping function is
enabled.
Check if this
requirement is satisfied
together with the IT
department of the
hospital.
3
Multicast
If the monitor adopts
multicast, the network
multicast function should
be enabled.
Check if this
requirement is satisfied
together with the IT
department of the
hospital.
4
Beacon and
DTIM
AP DTIM = 1, Beacon =
100 milliseconds
Check if this
requirement is satisfied
together with the IT
department of the
hospital.
5
Service port Refer to the white paper
of the Mindray monitor
network. The monitor
requests opening some
TCP/UDP ports.
Check if this
requirement is satisfied
together with the IT
department of the
hospital.
A.5
Using the Monitor to Evaluate Network Coverage
To confirm the coverage effect, perform the coverage test in the area where the patient
often goes.
Observe the RSSI displayed on the monitor. Check whether any offline event occurs, and
Содержание VS8
Страница 1: ...VS9 VS8 Vital Signs Monitor Service Manual ...
Страница 2: ......
Страница 8: ...VI VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 36: ...3 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 62: ...4 26 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 116: ...6 46 VS 9 VS 8 Vital Signs Monitor Service Manual 6 4 7 8 E 4 4 Removing the Touchscreen Assembly Same as E 4 3 ...
Страница 140: ...8 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 155: ......
Страница 156: ...PN 046 020105 00 2 0 ...