VS 9/VS 8 Vital Signs Monitor Service Manual
6-21
6.3.9
(F) Plastic Main Support Assembly
6.3.9.1
(F-1) Removing the CO
2
Module
Unscrew the 4 ST3.3X8 tapping screws at the position shown in the figure to replace the
CO
2
module assembly.
6.3.10
(E) Front Cover Assembly
6.3.10.1
(E-1) Removing the SpO
2
Board
Unscrew the screws corresponding on the small SpO
2
board as shown in figure on the
main board to replace the corresponding SpO
2
board.
Small SpO
2
board on Nellcor
Small SpO
2
board on Masimo
(2 M2X4 stainless steel pan head screws) (1 M2X4 stainless steel pan head screw) (2 M3X6
pan head screws with washers)