VS 9/VS 8 Vital Signs Monitor Service Manual
5-3
5.6.3
Alarm Lamp Failures
Fault Symptom
Possible Cause
Troubleshooting
Alarm lamp off or
cannot be turned
off, or alarm lamp
displays
abnormally
Cable failure
1. Verify the cable connecting the alarm lamp
board to the main board is properly connected.
2. Verify the cables and connectors are not
damaged.
Alarm lamp
board failure
Replace the alarm lamp board.
The main board
is faulty
Replace the main board
5.6.4
Button and Knob Failures
Fault Symptom
Possible Cause
Troubleshooting
Buttons do not
work
Cable failure
Verify the cable connecting the button board
to the main board is properly connected.
Button board
failure
Replace the button board
Knob does not
work
Cable failure
1. Verify the connection cable between the
knob and the button board.
2. Verify the cable connecting the button
board to the main board is properly
connected.
Knob damaged
Replace the rotary encoder.
Button board
failure
Replace the button board
Cable failure
1. Verify the signal cable and backlight
cable connecting the display to the
main board are correctly connected.
The touchscreen
does not respond.
Touchscreen disabled
Check if there is a symbol
shown
above the key on the main menu. If
yes, press Main Menu for more than 3s
to enable the touchscreen.
Cable failure
1. Verify the cable connecting the
touchscreen to the touchscreen
control board is properly connected.
2. Verify the cable connecting the
touchscreen control board to the main
board is properly connected.
Touchscreen control
board defective
Replace the touchscreen control board.
Touch screen defective
Replace the touchscreen
The main board is faulty
Replace the main board
Содержание VS8
Страница 1: ...VS9 VS8 Vital Signs Monitor Service Manual ...
Страница 2: ......
Страница 8: ...VI VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 36: ...3 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 62: ...4 26 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 116: ...6 46 VS 9 VS 8 Vital Signs Monitor Service Manual 6 4 7 8 E 4 4 Removing the Touchscreen Assembly Same as E 4 3 ...
Страница 140: ...8 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 155: ......
Страница 156: ...PN 046 020105 00 2 0 ...