VS 9/VS 8 Vital Signs Monitor Service Manual
6-19
6.3.7.3
(D-3) Removing the NIBP Pipeline
Disconnect the three air pipes of the pump and valve shown by the arrows by hand to
replace the NIBP airway.
6.3.8
(G) Removing the Sheet Metal Main Support and Plastic Main
Support Assemblies
Unscrew the 10 ST3.3X8 pan head screws as shown in the figure to disassemble the
assembly into the plastic main support assembly (F) and sheet metal main support
assembly (G).