VS 9/VS 8 Vital Signs Monitor Service Manual
4-11
4. Open the valve to start flowing CO
2
and make sure that there is sufficient flow to
vent to atmosphere.
5. Check that the real-time CO
2
value is within±0.3% in the
Calibrate
menu.
4.5.4.3
CO
2
Module Calibration
Calibrate the CO
2
module if you find it is inaccurate.
Calibration Preparations
Test tools:
Steel gas cylinder with 3%-7%±0.3% CO
2
and balance gas N
2
(P/N:
0075-00-0033-01)
T-shape connector
Tubing
Follow this Procedure to Perform a Calibration:
1. Make sure that the sidestream CO
2
module has been warmed up or started up.
2. Check the airway for leakage and perform a leakage test as well to make sure that
the airway has no leakage.
3. Open
Calibrate
:
Main Menu
->
Maintenance
-> enter the password ->
Module
->
CO2
->
Calibrate
.
4. In the
Calibrate
menu, select
Zero
.
5. After the zero calibration succeeds, connect the monitor and the test tools as
follows:
Monitor
Gas cylinder
Relief valve
T-shape connector
Tubing
Содержание VS8
Страница 1: ...VS9 VS8 Vital Signs Monitor Service Manual ...
Страница 2: ......
Страница 8: ...VI VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 36: ...3 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 62: ...4 26 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 116: ...6 46 VS 9 VS 8 Vital Signs Monitor Service Manual 6 4 7 8 E 4 4 Removing the Touchscreen Assembly Same as E 4 3 ...
Страница 140: ...8 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 155: ......
Страница 156: ...PN 046 020105 00 2 0 ...