VS 9/VS 8 Vital Signs Monitor Service Manual
6-35
6.4.3.5
(C-4) Removing the AC-DC Power Board
Unplug the power management board cable socket 2 (you may replace the cable only),
unplug the AC cable socket 1 shown in the figure, and unscrew the two M3*6 pan head
screws 3 to replace the power module.
6.4.3.6
(C-5) Removing the Panel and SpO
2
Signal Cable
1. Unplug the NIBP air tube 1 on the panel and unplug the SpO
2
cable 2 (two cables for
the dual-SpO
2
model and one cable for the single-SpO
2
model, subject to the actual
configuration of the monitor). Replace the VS8 panel assembly or SpO
2
cable
separately.
2. Rotate the SpO
2
cable clockwise to remove it. The new SpO
2
cable must be
connected at the correct position.