VS 9/VS 8 Vital Signs Monitor Service Manual
4-21
4.8
Battery Check
This test mainly tests the battery status.
4.8.1
Test Preparations
None.
4.8.2
Procedure
4.8.2.1
Functionality test
1. Remove any battery installed in the monitor.
2. Verify that the monitor works properly when running on AC power.
3. Install the battery according to the procedures described in the Operator's Manual.
4. Remove the AC power cord and verify that the monitor still works properly.
4.8.2.2
Performance test
Refer to the related content in
Battery
section in the operator’s manual and verify the
operating time of the battery meets the product specification.