6-6 VS 9/VS 8 Vital Signs Monitor Service Manual
6.3.2
(A-2) Recorder Assembly/Door
6.3.2.1
When No Recorder is Used
Insert a flat-bladed screwdriver or tweezers (or other similar small tools) into the removal
hole on the recorder door as shown in the figure to remove the recorder door.
6.3.2.2
When Recorder is Used.
1. Use a Phillips screwdriver to unscrew the 2 M3*6 pan head screws with washers at
the position shown in the figure on the rear cover.
2. Pinch the two lugs on the inside of the recorder as shown in the figure and press in
the direction of the arrow to remove the recorder.