3-4 VS 9/VS 8 Vital Signs Monitor Service Manual
3.3.1.2
Installation of VS8's Battery
Rotate the metal ring on the battery door at the bottom of the monitor by hand. After the
ring is unscrewed, the battery door can be opened to install the battery. Note: If a
small-capacity battery is installed, a small-capacity battery bracket needs to be used. If a
large-capacity battery is required, remove the small-capacity battery bracket from the
battery compartment and install the large-capacity battery.
When installing the battery, insert the battery cable into the socket on the board as
shown in the figure. (There are two sockets for the large-capacity battery, and only one
socket A for the small-capacity batteries).
After the battery is installed, reinstall the battery door on the rear cover and tighten the
metal ring.
Small-capacity
battery bracket
Metal ring on
the battery
door
Battery
door