VS 9/VS 8 Vital Signs Monitor Service Manual
8-5
8.2.2.1
Connecting Monitor and CMS
NOTE
Before allocating configuration for the monitor, exporting logs or
upgrading software through the CMS, build a connection between the
monitor and CMS.
Method 1: Choose
Main Menu
->
Maintenance
->
Network Setup
, select
Transfer
Setup
, turn on
Select CMS
, and then select
Add Central Station
. Enter a
Server
Address
, for example, 196.76.0.129, and click
Network Test
. When three green
ticks appear, the connection is successful.
Method 2: Choose
Main Menu
->
Maintenance
->
Network Setup
, select
Device
Discover
, enter the IP address of the main server, for example, 196.76.0.129, and
then check whether the main server is connected.
8.2.2.2
Upgrading CMS Station
1. Start the CMS station program, and enter a password to access the server setting
screen.
2. In
Device Management
, find
Version Upgrade Management
, as shown in the
following figure, and click
Add Version
.
3. The
Select Version File
screen is displayed. Select the
Install File
and
Version File
.
The Install File indicates the tool for installing the software. The VS Install File is
LEGO_Installer.pkg and the Version File is the software installation package. After the
selection, click
OK
.
Содержание VS8
Страница 1: ...VS9 VS8 Vital Signs Monitor Service Manual ...
Страница 2: ......
Страница 8: ...VI VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 36: ...3 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 62: ...4 26 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 116: ...6 46 VS 9 VS 8 Vital Signs Monitor Service Manual 6 4 7 8 E 4 4 Removing the Touchscreen Assembly Same as E 4 3 ...
Страница 140: ...8 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 155: ......
Страница 156: ...PN 046 020105 00 2 0 ...