VS 9/VS 8 Vital Signs Monitor Service Manual
A-13
Symptom
Possible Cause
Recommended Measures
meets requirements.
The SSID, IP address
obtaining mode and
safe mode are not
properly configured on
the monitor.
Reconfigure the information as
instructed in this manual.
The monitor fails.
Check if another monitor can be
connected online. If yes, restart this
monitor and ensure that the two
monitors have the same configuration. If
this monitor still cannot be connected
online, return it to Mindray for repair.
The monitor can
be connected to
the AP, but it
cannot be
connected to the
central
monitoring
system.
The monitor has not
obtained the access
permission of the
central monitoring
system.
Allow the monitor to access the central
monitoring system.
The monitor cannot
obtain any IP address,
and the IP addresses in
the IP address pool on
the DHCP server have
been used up.
Connect another network device to the
central monitoring system, and check if
the IP address can be obtained.
If the problem cannot be solved, contact
the IT department.
Static IP addresses
conflict.
Check if the monitor displays a message
about an IP address conflict.
If yes, ensure that all network devices
have unique IP addresses.
The network link fails.
Check if the center monitoring system
can be pinged through after the PC or
mobile phone is connected to the AP.
If the problem cannot be solved, contact
the IT department.
The service port
required by the monitor
is not enabled on the
hospital network.
Check if the service port required by the
monitor has been enabled on the
hospital network. If no, enable related
services (such as some UDP ports and
multicast service).
If the problem cannot be solved, contact
the IT department.
A single monitor
is disconnected
intermittently.
The monitor is moved
outside the coverage
area.
Find out the position of disconnection
and check whether the Wi-Fi at this
position is greater than
–
65 dBm.
The monitor fails.
Check if the monitor is easily
disconnected at the same position. If this
problem persists after the monitor is
Содержание VS8
Страница 1: ...VS9 VS8 Vital Signs Monitor Service Manual ...
Страница 2: ......
Страница 8: ...VI VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 36: ...3 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 62: ...4 26 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 116: ...6 46 VS 9 VS 8 Vital Signs Monitor Service Manual 6 4 7 8 E 4 4 Removing the Touchscreen Assembly Same as E 4 3 ...
Страница 140: ...8 12 VS 9 VS 8 Vital Signs Monitor Service Manual FOR YOUR NOTES ...
Страница 155: ......
Страница 156: ...PN 046 020105 00 2 0 ...