9
KHA3000
215
Ex
ecuting V
olta
g
e Fluctuation T
ests
Setting Test Conditions (61000-3-11)
Setting the Limit Value Standard
1
Press the Vf key.
The Vf-Test Conditions List display appears.
2
Set test conditions.
To use the same conditions as for a test that was executed in the past, call up a test
condition file and use it.
3
Press the F6 key to select menu page 1/4.
4
Press the F1 key to select Limitation Standard.
Select IEC 61000-3-11 Ed2.0.
5
Press the VIEW key.
The Vf-Observation and Analysis display appears.
Optimizing the Current Range
1
Press the VIEW key.
The Vf-Observation and Analysis display appears.
2
Press the F1 key to select List in View Type.
Measured values are displayed in the form of a list.
3
Press the VIEW key.
The Vf-Observation and Test Conditions display appears. The menu for setting the
measured values, which are still listed, is displayed.
4
Change the operating conditions of the EUT to maximize the input
current.
5
Press the F6 key to select menu page 4/4.
6
Press the F1 key to select V/I Range Setting.
A sub-menu is displayed.
7
Press the F1 key to select Ch. Setting (V/I Range).
When "Linked" is selected in the Channel Linkage menu, independent setting of L1, L2,
and L3 cannot be made.
8
Press the F2 key to select Current Range.
Select the range on the basis of the maximum current obtained in step 4. We
recommend performing the test using a fixed range.
If you do not know the current, select AUTO (auto range). In auto range, however, it
takes approximately 6 seconds for the measured values to stabilize after the range
switches. If a current variation causes the range to switch, wait at least 6 seconds
before starting the test. During the test, the range is fixed to the present range.
9
Press the F6 key to return.
S
ee
S
ee
S
ee
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...