6
KHA3000
135
Executing Harm
onic Tests
Setting Test Conditions (61000-3-12)
Set the test conditions with the HA-Test Conditions List display. When the test condition setting
is complete, switch to the HA-Observation and Analysis display for observation during the test.
Setting Limit Value Standard, Measuring Technique
Standard and Equipment
1
Press the HA key.
The HA-Test Conditions List display appears.
2
Select test conditions.
To use the same conditions as for a test that was executed in the past, call up a test
condition file and use it.
3
Press the F6 key to select menu page 1/4.
4
Press the F1 key to select Limitation Standard.
Select IEC 61000-3-12 2004 or IEC 61000-3-12 2011.
5
Press the F2 key to select Measuring Technique Standard.
To employ a measuring method for harmonic groups, select IEC 61000-4-7 Ed2.0 or
IEC 61000-4-7 Ed2.1.
6
Press the F3 key to select Equipment.
Select the equipment according to the type of the EUT.
Single-phase equipment includes the single-phase part of hybrid equipment. Balanced
three-phase equipment includes the three-phase part of hybrid equipment.
7
Press the F6 key to switch to another menu page and set other test
conditions.
Optimizing the Current Range
1
Press the VIEW key.
The HA-Observation and Analysis display appears.
2
Press the F1 key to select List in View Type.
Measured values are displayed in the form of a list.
3
Press the VIEW key.
The HA-Observation and Test Conditions display appears and the menu for setting the
measured values, which are still listed, is displayed.
4
Change the operating conditions of the EUT to maximize the input
current.
5
Press the F6 key to select menu page 4/4.
See
See
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......