5
KHA3000
99
61000-3-2
Setting the Harm
onic Test
7
Limit Values
(IEC standard)
(Class C)
Select the limit value.
Normal limit
value
Normal
The class C limit value is specified. Select this item for
lighting equipment exceeding 25 W.
Class A
Select this item for incandescent lighting equipment with a
dimmer exceeding 25 W.
Class D
Select this item for discharge lighting equipment of up to
25 W.
3rd/5th/
Current
Wave
Select this item for discharge lighting equipment of 25 W or
lower. Only selectable for IEC 61000-3-2 Ed5.0 Class C.
This is only valid for single-phase, two-wire circuits.
If you select this item, you will not be able to print test result
reports.
THD
Limit
Select this item for discharge lighting equipment of 5 W or
grater and 25 W or lower. Only selectable for IEC 61000-3-2
Ed5.0 Class C.
Limit Values
(JIS standard)
(Class C) Select the limit value.
Normal limit
value
Normal
The class C limit value is specified. Select this item for
lighting equipment exceeding 25 W (35 W for household
luminaire).
Class A
Select this item for incandescent lighting equipment with a
dimmer exceeding 25 W (35 W for household luminaire).
Class D
Select this item for discharge lighting equipment of up to
25 W and compact fluorescent lamps exceeding 25 W and
of up to 35 W.
3rd/5th/
Current
Wave
Only selectable for JIS C61000-3-2 2019 Class C.
This is only valid for single-phase, two-wire circuits.
If you select this item, you will not be able to print test result
reports.
THD
Limit
Select this item for discharge lighting equipment of 5 W or
grater and 25 W or lower. Only selectable for IEC 61000-3-2
Ed5.0 Class C.
8
Return
Terminates the class option selection.
–
9
600 W Air Conditioner
(JIS class A) Specify whether air conditioners have an effective power
exceeding 600 W.
No
10
Prev/Next Page 3/4
Switches to another menu page.
–
F key: Function key. You can also use the ESC key to return.
If you want to use the values (specific values) specified by the manufacturer in place of the
actually measured values for the power, fundamental current, and power factor of the EUT,
select "Measured" to execute the test. Then, check that the specified value is within ±10 % of
the measured value, and execute the test using specific values.
Write the test results using measured values (power (Class D), fundamental current, and
power factor (Class C)) in the Memo column of the report for tests using specific values, or
prepare each report separately and use them together.
Testing is possible even when the difference between measured values and specific values
exceeds ±10 %, but it is important to note that the limit value of the obtained test results does
not meet the standard.
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......