6
KHA3000
137
Setting Test Conditions (61000-3-12)
Executing Harm
onic Tests
7
Press the F6 key to return.
8
Press the F6 key to select menu page 3/4.
9
Press the F1 key to select Specified in Reference Current/Reference
Fundamental Current.
10
Press the F2 key to input a reference current specigication/reference
fundamental current specification.
The input range is 0.0 A to 75.0 A.
Inputting Comments for Report Printing
1
Press the HA key.
The HA-Test Conditions List display appears.
2
Press the F6 key to select menu page 2/4.
3
Press the F3 key to select Comment Input.
A sub-menu is displayed.
4
Input comments for the menu items.
The input comments can be checked on the screen.
5
Press the F6 key to return.
Checking the Third, Fifth and Seventh Harmonics
1
Press the VIEW key.
The HA-Observation and Analysis display appears.
2
Press the F6 key to select menu page 1/2.
3
Press the F1 key to select Harmonics List in View Type.
The harmonics list of measured values is displayed.
4
Press the F2 key to select View Setting.
A sub-menu is displayed.
5
Press the F4 key to select In/Iref [%] or In/I1 [%] in Measurement Value
Selection.
6
Press the F5 key to select L1 in Check View Phase.
Select the L1 phase first. Make checks for each phase by the following procedure.
7
Check that the largest third harmonic current value is less than 5% of the
reference current value/reference fundamental current value.
Check for each phase.
See
See
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......