Messages
372
KHA3000
SOURce subsystem
STATus subsystem
IEEE488.2 common commands
SCPI command
Setting
Default
Response Influence
Description
R/W
Note
Program header
Parameter
Unit
SOUR
:OUTP[:STAT]
bool
ON | OFF
–
OFF
NR1
ON/OFF of the output of the AC
power supply
R/W
3
:PHAS:ON
NRf
0 to 359, -1
DEG
–
NR3
Output ON phase of the AC power
supply
R/W
3
:FREQ
NRf
50, 60
HZ
–
NR3
Frequency of the AC power supply R/W
3
:VOLT
NRf
0.0 to 305.0
V
–
NR3
Output voltage of the AC power
supply
R/W
3
:VOLT:RANG
NRf
100 | 200
V
–
NR3
Output voltage range of the AC
power supply
R/W
3
SCPI command
Setting
Default
Response Influence
Description
R/W
Note
Program header
Parameter
Unit
STAT
:OPER
[:EVEN]
–
–
–
–
NR1
–
Queries an event.
*1
R
1
:COND
–
–
–
–
NR1
–
Queries the register status.
*1
R
1
:ENAB
NRf
0 to 32767
–
–
NR1
–
Enable
*1
R/W
1
:NTR
NRf
0 to 32767
–
–
NR1
–
Negative transition
*1
R/W
1
:PTR
NRf
0 to 32767
–
–
NR1
–
Positive transition
*1
R/W
1
:PRES
–
–
–
Initializes the filter register.
W
1
:QUES
[:EVEN]
–
–
–
–
NR1
–
Queries an event.
*2
R
1
:COND
–
–
–
–
NR1
–
Queries the register status.
*2
R
1
:ENAB
NRf
0 to 32767
–
–
NR1
–
Enable
*2
R/W
1
:NTR
NRf
0 to 32767
–
–
NR1
–
Positive transition
*2
R/W
1
:PTR
NRf
0 to 32767
–
–
NR1
–
Negative transition
*2
R/W
1
*1.
OPERation status register
*2.
QUEStionable status register
IEEE488.2
common
command
Parameter
Description
R/W
*CLS
–
Clears all event registers.
W
*ESE
NR1
Sets the event status enable register bit.
R/W
*ESR
–
Queries the event status register.
R
*IDN
–
Queries the identification string (information on the manufacturer).
R
*OPC
–
When all operations of the equipment that were detected as in standby are finished, the
equipment generates an operation completion message in the event status register.
R/W
*OPT
–
Queries the options installed in this product.
R
*RST
–
Resets the equipment to the default state independent of the use history of the equipment.
W
*SRE
NR1
Sets the service request enable register.
R/W
*STB
–
Queries the contents of the status byte register and master summary status message.
R
*TRG
–
Trigger command
W
*TST
–
Executes self-diagnosis.
R
*WAI
–
Prevents the equipment from executing any command or query until all operations in
standby are complete.
W
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......