108
KHA3000
61000-3-2
Adjusting the phase difference between voltage and current
When using external current sensors that are connected to the EXT INPUT (BNC) terminals
on the rear panel, the phase difference between the current detected by the current sensor
and the voltage measured by this product at a voltage measurement point can be adjusted.
For the phase adjustment, pure resistors are connected in place of the EUT.
The adjustment is made by observing the power factor shown on the screen of this product.
The adjustment is made separately for each phase. Incorrect adjustment can adversely affect
the measurement of real power.
The components to be prepared by the customer
●
Resistance Load
Non-inductive resistors are required. Be sure to use resistors of which power rating is as
close as to the power consumption of the EUT.
Connecting Resistance Loads
Connect pure resistors in place of the EUT.
1
Turn off the output of the AC power supply.
2
Connect the resistance load(s) in place of the EUT.
3
Turn on the output of the AC power supply.
Set the output voltage to the rated voltage of the EUT.
Setting of Observation Display (HA- Observation/Analysis
Display)
Observe the power factor value on the list display.
1
Press the VIEW key.
The HA-Observation/Analysis display will open.
2
Press the F6 key to select Page 1/2 of the menu.
3
Press the F1 key for the list display type.
A list of measured values will be presented.
4
Press the F1 key to select View Items.
A sub-menu is displayed.
5
Turn the small knob to select Power Factor.
6
Press the ENTER key to save the selection.
The Power Factor check box will be checked. You can freely select other view items at
the same time.
WARNING
•
This work has a risk of electric shock. For the AC power supply used for the
measurement circuit, be sure to disconnect the power cord from the outlet, turn off
the switch on the switchboard to which the power cord is connected, and turn off the
POWER switch.
See
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......