Specifications
Specifications
Harmonic Voltage Measurement Function (Measurement
Power Supply Quality Check Function)
Flicker/voltage Fluctuation Analysis Function
Item
Specifications
Measurement item
Voltage, frequency, voltage harmonic content
Voltage harmonic analysis order
40th order
Voltage harmonic
measurement
accuracy
*1
(TYP value)
*1.
Fundamental frequency: 45 Hz to 65 Hz; PLL is locked in the effective input range. Specified for the am-
bient temperature of 18
°C
to 28
°C
and relative humidity of 20 % to 80 % (without condensation). When
scaling is used, the voltage range shown on the display is the "range" value.
150 V range
±
((0.4+0.417
×
n kHz)% of rdng+0.04 % of range+1digit)
: n is frequency.
300 V and 600 V
range
±
((0.4+0.417
×
n kHz)% of rdng+0.04 % of range+2digits)
: n is frequency.
Item
Specifications
Conforming standard
IEC 61000-3-3 Ed3.1(2017), IEC 61000-3-3 Ed2.0(2008)
IEC 61000-3-11 Ed2.0(2007)
Requirements for measuring
instrument
*1
*1.
Acceptable if it is within the specified range.
IEC 61000-4-15 Ed2.0(2010), IEC 61000-4-15 Ed1.1(2003)
Flicker
Pst accuracy
*2
*2.
At input of IEC 61000-4-15 Table 1 signal.
1
±
0.05
Pst measurement time 30 to 900 seconds
Plt measurement
count
1 to 12
Voltage
fluctuation
Steady voltage
recognition condition
Steady state recognized when fluctuation width
±
0.1 % or less for
1 second or more
Measurement time/
count
Conforming to the Pst measurement time and Plt measurement
count
Measurement
method
*3
*3.
d measurement in independent mode applies only to dmax measurement of manual switching equipment.
Together with Pst or independent mode can be selected
dmax
measurement
of manual
switching
equipment
Measurement count
3 to 24 times
Time of one
measurement session
30 to 180 seconds
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......