Setting the Voltage Fluctuation Test
61000-3-3
Limitation Standards and Nominal Values
●
Pst Auto
Simultaneously executes the d measurement (dmax, Tmax (or d(t)>3.3%), dc) and the Pst
and Pit (flicker) measurement. The maximum value for each Pst measurement segment
time is displayed as the results of the d measurement.
●
Manual
Executes measurement with a method that conforms to
“
Test conditions and procedures
for measuring voltage changes caused by manual switching.
”
This product makes
judgment by using the arithmetic average of 22 measured values obtained by excluding
the maximum and minimum values of 24 collected measurements.
No.
Menu item
Description
Default
1
Limitation Std
Select a limit value standard. Select either
“
Pst Auto
”
or
“
Manual
”
for the
measuring method for IEC 61000-3-3. When IEC 61000-3-11 Ed2.0 is
selected, the menu for equipment with rated current of 75 A or less is
displayed. The limitation standard IEC 61000-3-3 Ed to be set depends
on the selection of the standard for measurement techniques.
Abbreviated Standard names are shown on the display.
IEC 61000-3-3
(Pst Auto)
2
Meas Technic
Select a measurement technique. If IEC 61000-4-15:Ed2.0 is selected,
the limitation standard is set to IEC 61000-3-3:Ed3.1(2017). If IEC
61000-4-15:Ed1.1 is selected, the limitation standard is set to IEC
61000-3-3:Ed2.0(2008). Abbreviated Standard names are shown on the
display.
IEC 61000-4-15
Ed2.0
3
Nominal Voltage
Specify the rated power voltage of the EUT. Select
“
230 V
”
(fixed value)
or
“
Specified.
”
230 V
4
Specified Nominal Volt
When
“
Specified
”
was selected in the F1 menu, enter the rated power
voltage of the EUT.
The input range is 100 V to 600 V.
Enter the value with the numeric keypad and press the ENTER key, or
use the large knob to increase or decrease the value.
100 V
5
Nominal Frequency
Select the rated power frequency of the EUT.
50 Hz
6
Return
Terminates the specification of the nominal voltage and frequency.
–
7
To Meas Setting Page
Switches to page 4/4.
–
8
Prev/Next
Page 1/4
Switches to another menu page.
–
F key: Function key. You can also use the ESC key to return.
Menu 1/4
F1
Limitation Std
IEC 61000-3-3 (Pst Auto)
IEC 61000-3-3 (Manual)
IEC 61000-3-11 Ed2.0
F2
Meas Technic
IEC 61000-4-15 Ed1.1
IEC 61000-4-15 Ed2.0
F3
Nominal Values
Sub Menu
F5
F6
F1
Nominal Voltage
230 V
Specified
F2
Specified Nominal Volt
(Input numeric value)
F3
Nominal Frequency
50 Hz
60 Hz
F6
Return
To Meas Setting Page
→
1
3
2
4
5
6
7
8
Prev/Next
Page 1/4
IEC 3-3 Auto
IEC 3-3 Manu
JIS 3-11 Ed2.0
Display of abbreviated
Limitation Standard names
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......