256
KHA3000
Basic Measurement
15
_ V/I Phase Adj. (deg)
Adjust the current phase of the external current sensor. This item is
enabled when "BNC" is selected in the Current Input Terminal menu. It is
independent of the Enable/Disable setting of the Current Scaling menu.
The input range is -180.00 to +180.00 degrees.
Use the ten-key to enter the value and press the ENTER key or use the
large knob to increment or decrement the value.
The input phase (channel L1, L2 or L3) that has been selected in Ch Setting
(Scaling) will be shown at the underlined position of the menu item.
0.00
16
All-Ch lump Set. Exec.
Copy the current channel settings set for execution to all other input
channels (channels L1, L2, and L3) for setting them identically. When the
F key is pressed, a confirmation dialog will be displayed. If you are sure to
copy them, press the F1 key, "OK". If not, press the F2 key, "Cancel".
* Whenever "Sensor" is selected for Current Input Terminal, copy the
Current Input Terminal settings to all other input channels (channels L1,
L2, and L3) for setting them identically. Items such as CT Ratio will not
be copied.
-
17
Prev/Next Page 1/4
Move to other sub-menu pages.
-
18
_Volt. Scaling
Enable or disable the scaling of the voltage ratio for the external PT
(potential transformer). If "Enable" is selected, the SC icon will be shown
in the V RANGE part of the display.
The input phase (channel L1, L2 or L3) that has been selected in Ch Setting
(Scaling) will be shown at the underlined position of the menu item.
-
19
_ PT Ratio
This item is enabled when "Enable" is selected in the Volt. Scaling
menu. Here, the voltage ratio (scaling) of PT (potential transformer) is
set. The input range is 0.001 to 100.000. Use the ten-key to enter the
value and press the ENTER key or use the large knob to increment or
decrement the value.
The input phase (channel L1, L2 or L3) that has been selected in Ch Setting
(Scaling) will be shown at the underlined position of the menu item.
1.000
20
_ Volt. CF
This item is enabled when "Enable" is selected in the Volt. Scaling
menu. In order to measure the peak value without the effect of any
distortion, the peak to rms ratio (CF) is set for the range currently set.
The CF setting affects the measurement resolution.
The input range is 1.00 to 2.00.
Use the ten-key to enter the value and press the ENTER key or use the
large knob to increment or decrement the value. The input phase
(channel L1, L2 or L3) that has been selected in Ch Setting (Scaling) will
be shown at the underlined position of the menu item.
2.00
21
Prev/Next Page 2/4
Move to other sub-menu pages.
-
22
_ Curr. Scaling
Enable or disable the scaling of the current signal from the external
current sensor. If "Enable" is selected, the SC icon will be shown in the I
RANGE part of the display.
When "Sensor" is selected in the Current Input Terminal menu, this item
is fixed and grayed out in the "Enabled" status. The input phase (channel
L1, L2 or L3) that has been selected in Ch Setting (Scaling) will be
shown at the underlined position of the menu item.
-
23
_ CT Ratio
This item is enabled when "Enable" is selected in the Current Scaling
menu. This item is enabled when "Shunt" is selected in the Current Input
Terminal menu. Here, the scaling ratio of CT (current transformer) is set.
The input range is 0.001 to 1000.000.
Use the ten-key to enter the value and press the ENTER key or use the
large knob to increment or decrement the value.
The input phase (channel L1, L2 or L3) that has been selected in Ch Setting
(Scaling) will be shown at the underlined position of the menu item.
1.000
24
_ Curr. CF
This item is enabled when "Enable" is selected in the Current Scaling
menu. This item is enabled when "Shunt" or "BNC" is selected in the
Current Input Terminal menu. In order to measure the peak value without
the effect of any distortion, the peak to rms ratio (CF) is set for the range
currently set. The CF setting affects the measurement resolution.
The input range is 1.00 to 4.00.
Use the ten-key to enter the value and press the ENTER key or use the
large knob to increment or decrement the value. The input phase
(channel L1, L2 or L3) that has been selected in Ch Setting (Scaling) will
be shown at the underlined position of the menu item.
4.00
See
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......