Common Information on Measurement Circuits
W
iring M
easurem
ent Circuits
Parts to be Prepared by the Customer
●
Electric wire
Used to connect the devices of a measurement circuit
●
Crimping terminal
Attached to the end of an electric wire to securely connect the wire to the terminal of a
device
●
Relay outlet
Used when the EUT uses a power cord with a plug
●
The following four types of impedance are used according to limit value standards
IEC 61000-3-11 and IEC 61000-3-12:
•
0.25
+ jn0.25
(single-phase)
•
0.15
+ jn0.15
(three-phase)
•
0.10
+ jn0.10
(three-phase neutral wire)
•
The voltage drop caused by the EUT falls within the range from 3 % to 5 % of the voltage
and the ratio of the reactance component to the resistance component falls within the
range from 0.5 to 0.75.
Selecting the Type of Electric Wire
Select the type of wire according to the input current of the EUT. For a rough idea for
selection, the table below lists nominal sectional areas of wire and the corresponding
permissible current values.
Use a crimping terminal to terminate the wire. The crimping terminal must have a screw
mounting hole with a diameter of 6 mm and fit the wire. When the EUT uses a power cord
with a plug, connect the wire to the LOAD terminals using a relay outlet. Do the same for the
wire used for the relay outlet.
L
N
G
Nominal sectional
area [mm
2
]
AWG
(Reference
sectional area)
[mm
2
]
Permissible current
*1
[A]
(Ta = 30
°C
)
*1.
Excerpt from Japanese laws related to electrical equipment
Current
recommended by
Kikusui [A]
2
14
(2.08)
27
10
3.5
12
(3.31)
37
-
5.5
10
(5.26)
49
20
8
8
(8.37)
61
30
14
6
(13.3)
88
50
•
Kikusui recommends wire of a nominal sectional area of 8 mm
2
or larger. Use as short a wire
as possible. Kikusui recommends a wire length of 1.5 m or shorter. For the test system, note
the IEC 61000-4-7 standard requirements (voltage drop by wiring impedance).
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......