When the Test Time Expires (61000-3-12)
Executing Harm
onic Tests
4
Press the START key.
A confirmation dialog will be presented regarding the line impedance, optional current
sensor or any negative sum (
) of the real power.
5
After checking presence/absence of impedance and the setting of line
impedance, select the F1 key (OK).
If the F2 key (Cancel) is selected, presence/absence of impedance is neglected. To
start the test, press the START key again.
The test is started.
6
When the test ends, execute Step 1 in "When the Test Time Expires
(61000-3-12)" on page 143.
7
If the general judgment is FAIL, return to Step 1.
Repeat the procedure until the general judgment becomes PASS.
When the general judgment becomes PASS
8
Terminate the test.
Select an item that you want to operate in the test ending menu. For details on the
selection, see "Overview of the Test Ending Menu" on page 146.
9
Compare the set judgment Rsce and Rsce (estimated value) with each
other.
When the judgment Rsce is greater than 33 and equal to or
smaller than the Rsce (estimated value)
10
Calculate the Ssc from the judgment Rsce and rated apparent power of
the equipment.
For details on calculating the Ssc, see "Ssc calculation" on page 144.
When the judgment Rsce is greater than 33 and exceeds the
Rsce (estimated value)
11
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......