278
KHA3000
Using the RS232C Interface
The RS232C port on this product is a standard D-sub 9-pin male connector.
Check that the POWER switches of this product and computer are off, and connect this
product to the computer using a standard cross cable (null modem cable).
Use a D-sub 9-pin female-to-female AT type for the cross cable. The connector pin
arrangement is shown below.
This product does not use hardware handshaking (as shown in the cross cable example 2).
RS232C setting
1
Press the SYSTEM key.
The System Setup display appears.
2
Press the F2 key (I/F Setting
Sub Menu).
A sub-menu is displayed.
3
Press the F1 key (I/F Select) a few times or turn the small knob to select
RS232C.
4
Press the F3 key (RS232C baudrate).
5
Press the F3 key (RS232C baudrate) a few times or turn the small knob to
set the baudrate.
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
D-sub 9-pin female
D-sub 9-pin female
Cross cable example 1
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
D-sub 9-pin female
D-sub 9-pin female
Cross cable example 2
#4-40UNC
Inch screw
#4-40UNC
Inch screw
1: CD (carrier detect)
2: RXD (receive data)
3: TXD (transmit data)
4: DTR (data terminal ready)
5: GND (signal ground)
6: DSR (data set ready)
7: RTS (ready to send)
8: CTS (clear to send)
9: RI (ring indicator)
Facing the KHA3000 rear panel
See
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......