6
KHA3000
123
Executing Harm
onic Tests
Setting the LIN and AC Power Supply (61000-3-2)
Set the LIN (Line Impedance Network) and AC power supply before turning on the power to
the EUT.
1
Set the impedance of the LIN.
Operate the LIN panel to set the impedance to OUT (THRU).
2
Press the EXT CONT key.
The External Control display appears.
3
Press the F1 key to select AC Power Supply Control.
A sub-menu is displayed.
4
Press the F2 and F3 keys to set the output voltage and frequency
respectively.
Set the AC power supply voltage and frequency according to the rated power of the EUT.
5
Press the F1 key to set OUTPUT to ON.
OUTPUT of the AC power supply is set to ON. The lamp icon in the upper part of the
screen lights up.
When it is synchronized with the frequency of the AC power supply, the triangular wave
in the PLL icon in the upper part of the screen becomes still.
6
Press the F6 key to return.
7
If you have set the limitation standard to JIS 61000-3-2 2005 or JIS 61000-
3-2 2019, use the AC power source confirmation screen to record the
measured values.
Disconnect the EUT's AC power supply before you record the values.
The recorded values are held until the KHA3000 is turned off. You can use the SD006-
KHA application software to print these recorded values.
8
Turn on the power to the EUT.
See
See
If the triangular wave in the PLL icon does not become still
AC power supply
Check whether it is output correctly.
Wiring
Check whether the plug for the voltage-sensing terminal is set and
connected to the VOLTAGE SENSING terminal on the rear.
Frequency
The PLL lock frequency range is 45 Hz to 65 Hz.
See
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......