Messages
368
KHA3000
VFLuctuation subsystem
HARM
:IEC312
:ETYP
char
SING | W3P3 |
BAL | UNB
–
BAL
char
Type of the EUT
R/W
3
:FUND
:DEF
char
MEAS | SPEC
–
MEAS
char
Method of specifying the reference
fundamental current
R/W
3
:CURR
numeric
0.0 to 20.0
A
20.0
NR3
Specified value of reference
fundamental current
R/W
3
:JUDG:RSCE
NRf
33 to 350
–
33
NR1
Judgment Rsce
R/W
3
:LVT
char
UNB | BAL |
SPEC
–
UNB
char
Limit value to be applied
R/W
3
:NOM:SVOL
NRf
100 to 600
V
400
NR3
Normal system voltage
R/W
3
:RAT
:DEF
char
MEAS | SPEC
–
MEAS
char
Method of specifying the rated
current
R/W
3
:CURR
NRf
0.1 to 40.0
A
20.0
NR3
Rated current of the EUT
R/W
3
:VOLT
NRf
100 to 600
V
230
NR3
Rated voltage of the EUT
R/W
3
SCPI command
Setting
Default
Response Influence
Description
R/W
Note
Program header
Parameter
Unit
SCPI command
Setting
Default
Response Influence
Description
R/W
Note
Program header
Parameter
Unit
VFL
:STAN
:LIM
[:NAME]
string
“IEC
61000-3-
3(Pst AUTO)” |
“IEC
61000 3-
3(ManualSW)
”
|
“IEC 61000-3-
11 Ed2.0”
–
“IEC
61000-3-
3(Pst
AUTO)”
string
Limit value standard
R/W
3
:CAT
string
–
–
–
string
Queries available limit value
standards.
R
3
:ALI[:STAT
]
bool
ON | OFF
–
OFF
NR1
Specifies the naming of Standard
R/W
3
:ALI:NAM
E
string
–
–
Standar
d name
string
–
Limitation standard name
R/W
3
:MTEC
[:NAME]
string
“IEC
61000-4-
15 Ed1.1” |
“IEC 61000-4-
15 Ed2.0”
–
“IEC
61000-4-
15
Ed1.1”
string
Measuring technique standard
R/W
3
:CAT
string
–
–
–
string
Queries available measuring
technique standards.
R
3
:ALI:NAM
E
string
–
–
Standar
d name
string
–
Limitation standard name
R/W
3
:NOM
:FREQ
NRf
50 | 60
HZ
50
NR3
Nominal frequency
R/W
3
:VOL
NRf
100 to 600
V
230
NR3
Nominal voltage
R/W
3
:TEXT
:COMM
encoded-
string
0
“
Memo,
”
etc.
–
All
items“ ”
string
Comment text
R/W
3
:IEC33
:DMAX
NRf
4 | 6 | 7
PC
T
6
NR3
dmax limit value
R/W
3
:DCO
NRf
3 to 24
–
24
NR3
d measurement count
R/W
3
:DTIM
numeric
30 to 180
S
60
NR3
d measurement time
R/W
3
:PSTC
NRf
1 to 12
–
12
NR3
Pst measurement count
R/W
3
:PSTT
numeric
30 to 900
S
600
NR3
Pst measurement time
R/W
3
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......