9
KHA3000
207
Ex
ecuting V
olta
g
e Fluctuation T
ests
Setting the Impedance and AC Power Supply
The following procedure is common to the 61000-3-3 and 61000-3-11 standards.
1
Set the impedance.
See "Setting the Impedance" below.
2
Press the EXT CONT key.
The External Control display appears.
3
Press the F1 key to select AC Power Supply Control.
A sub-menu is displayed.
4
Press the F2 and F3 keys to set the output voltage and frequency
respectively.
Set the AC power supply voltage and frequency according to the rated power of the EUT.
5
Set OUTPUT of the AC power supply to ON.
OUTPUT of the AC power supply is set to ON. The lamp icon in the upper part of the
screen lights up.
When it is synchronized with the frequency of the AC power supply, the triangular wave
in the PLL icon in the upper part of the screen becomes still.
6
Press the F6 key to return.
7
Turn on the power to the EUT.
Setting the Impedance
61000-3-3
Set to the reference impedance. The Line Impedance Network can be used; for details on
how to select the impedance, refer to the Operation Manual of the Line Impedance Network.
•
0.40
+ jn0.25
(single-phase)
•
0.24
+ jn0.15
(three-phase)
•
0.16
+ jn0.10
(three-phase neutral line)
61000-3-11
Customers are asked to prepare the impedance.
●
When you declare that the power supply to be connected requires a current
capacity of 100 A or more per phase as the operating condition of the equipment
•
0.25
+ jn0.25
(single-phase)
•
0.15
+ jn0.15
(three-phase)
•
0.10
+ jn0.10
(three-phase neutral line)
●
When you declare the maximum permissible system impedance (Zmax) of the
power supply to be connected as the operating condition of the equipment
•
The voltage drop caused by the EUT is 3% to 5% of the voltage.
•
The ratio of the reactance component to the resistance component of the impedance is
0.5 to 0.75.
See
If the triangular wave in the PLL icon does not become still
AC power supply
Check whether it is output correctly.
Wiring
Check whether the plug for the voltage-sensing terminal is set and
connected to the VOLTAGE SENSING terminal on the rear.
Frequency
The PLL lock frequency range is 45 Hz to 65 Hz.
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......