172
KHA3000
View Setting Sub-menu
●
61000-3-12 2004
When In/I1[%] is selected with the F4 key
Measured values are those in a selected data frame. They are expressed as the ratio to the
reference fundamental current.
●
61000-3-12 2004
When In[Arms] is selected with the F4 key
Measured values are those in a selected data frame.
Item
Order
Lim1[%]
Meas[%]
Angle[%]
Ave[%]
Per[%]
Max[%]
Description
Harmonic
orders 1st to
40th.
The 1st
order is the
reference
fundamental
current,
THD and
PWHD.
Limit value (%)
Measured
value of
harmonic
current
Ratio (%) to
reference
fundamental
current
Measured
value of
phase angle
(Upper)
Average value (%)
from the beginning
of the test
When
“
Average
”
is
selected with the
F1 key
Ratio of the
Ave[%] to limit
value when
“
Average
”
is
selected with the
F1 key
Ratio of the
Max[%] to limit
value when
“
Maximum
”
is
selected with the
F1 key
(Lower)
Maximum value
(%) from the
beginning of the
test
When
“
Maximum
”
is selected with
the F1 key
“
-----
”
means there
is no numeric value
to be displayed:
•
When limit values
do not exist
•
When a
calculated limit
value is less than
0.0001 A (0.001
A in the 10 A
range or higher)
Display color
changes.
“
-----
”
means there
is no value to be
displayed:
•
In the test
condition setting
state (real-time
measurement)
Item
Order
Lim1[A]
Meas[Arms] Angle[deg]
Ave[Arms]
Per[%]
Max[Arms]
Description
Harmonic
orders 1st to
40th.
The 1st
order is the
reference
fundamental
current,
THD and
PWHD.
“
-----
”
means there
is no value to be
displayed.
Measured
value of
harmonic
current (A)
Measured
value of
phase angle
(Upper)
Average value (A)
from the beginning
of the test
When
“
Average
”
is
selected with the
F1 key
Ratio of the
Ave[%] to limit
value when
“
Average
”
is
selected with the
F1 key
Ratio of the
Max[%] to limit
value when
“
Maximum
”
is
selected with the
F1 key
(Lower)
Maximum value
(A) from the
beginning of the
test
When
“
Maximum
”
is selected with
the F1 key
“
-----
”
means there
is no numeric value
to be displayed:
•
When limit values
do not exist
•
When a
calculated limit
value is less than
0.0001 A (0.001
A in the 10 A
range or higher)
Display color
changes.
“
-----
”
means there
is no value to be
displayed:
•
In the test
condition setting
state (real-time
measurement)
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......