318
KHA3000
Voltage Fluctuation Measurement Mode (61000-3-11)
VFL:IEC331:MARG:FLIC
Sets the margin for the limit value of Pst or Plt. Set a relative value of the standard limit value
(100). The setting range is 10 to 100. For instance, specify 80 to set the margin to 80 % of the
standard limit value.
Command
VFLuctuation:IEC33:MARGin:FLICker {<numeric>|MIN|MAX}
VFLuctuation:IEC33:MARGin:FLICker?
Parameter Value:
10 to 100 (The default is 100.)
Unit:
PCT
Response
Returns the margin for the limit value of Pst or Plt in the <NR3> format.
Impedance
VFL:IEC311:TIMP
Sets the specified value of the test impedance. Setting a value cancels the default value.
Command
VFLuctuation:IEC311:TIMPedance[:VALues]
<numeric>,<numeric>,<numeric>,<numeric>
VFLuctuation:IEC311:TIMPedance[:VALues]?
Parameter Value:
R
A
,X
A
,R
N
,X
N
: 0.00 to 1.00
(The default of R
A
and X
A
is 0.24 and 0.15 and that of R
N
and X
N
is 0.16 and
0.10.)
Unit:
OHM
Response
Returns the specified value of the test impedance in the <NR3> format.
VFL:IEC311:TIMP:DEF
Sets the definition of the test impedance. As the definition, a fixed standard value or an
optional value that can be specified by the user can be used.
Command
VFLuctuation:IEC311:TIMPedance:DEFinition
{STANdard1|STANdard3|SPECified1|SPECified3}
VFLuctuation:IEC311:TIMPedance:DEFinition?
Parameter Value:
STANdard1
Standard value (single-phase)
STANdard3
Standard value (three-phase)
SPECified1
Specified value (single-phase)
SPECified3
Specified value (three-phase) (default)
Response
Returns the type of the test impedance value in the <character> format.
** RST
RST
** RST
RST
** RST
RST
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......