Controlling the AC Power Supply
Setting the Harm
onic Test
Voltage and Frequency Settings and ON Phase
No.
Menu item
Description
Default
1
OUTPUT
Turn on or off the output of the AC power. In the ON state, the icon in the
upper part of the screen lights up.
OFF
2
Voltage Setting
Enter the output voltage of the AC power. The input range is 0.0 V to
304.8 V.
Specify the voltage according to the rated voltage of the EUT. This value
determines the voltage range that will be decided in the F4 menu item.
The type of the set value of the output voltage depends on the wiring
method. If the wiring method is single-phase two-wire, single-phase
three-wire, or three-phase four-wire, the phase voltage is used. In the
case of three-phase three-wire, the line voltage is used. (The menus for
setting the wiring method are described in the chapters of harmonic
test, voltage fluctuation test and other tests.)
Enter the value with the numeric keypad and press the ENTER key, or
use the large knob to increase or decrease the value.
–
3
Freq Setting
Select the output frequency of the AC power according to the rated
frequency of the EUT.
50 Hz
4
Voltage Range
Select the output voltage range of the AC power.
Select the 100 V range for an output voltage setting range of 0 V to
152.5 V.
Select the 200 V range for an output voltage setting range of 0 V to
304.8 V.
100 V
5
ON Phase
The voltage phase angle when the output of the AC power is turned on
can be specified. In this menu item, select a specification method.
When
“
Free
”
is selected, the voltage phase is not specified. When
“
Specified
”
is selected, enter a value in the F2 menu item, Phase Angle.
Free
6
Phase Angle
Enter a value of the phase angle when
“
Specified
”
is selected in the F1
menu item. The input range is 0° to 360°.
Enter the value with the numeric keypad and press the ENTER key, or
use the large knob to increase or decrease the value.
0
7
Return
Terminates the ON phase selection.
–
8
Return
Terminates the control of the ON phase.
–
9
Alarm Clear
Clears the alarm of the AC power.
–
10
AC Pow Source Model
Select
Don’t use or Use
.
Don
’
t use
F key: Function key. You can also use the ESC key to return.
F1
AC Power Supply Cont
Sub Menu
F1
OUTPUT
OFF
ON
F2
Voltage Setting
(Input numeric value)
F3
Freq Setting
50 Hz
60 Hz
F4
Voltage Range
100 V
200 V
F5
F1
On Phase
Free
Specified
F2
Phase Angle
(Input numeric value)
F6
Return
F6
Return
F3
Alarm Clear
F4
AC Pow Source Model
Don’t use
Use
On Phase Select
Æ
Sub Menu
1
3
2
4
5
6
7
8
9
10
Menu 1/1
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......