KHA3000
View Setting Sub-menu
List
No.
Menu item
Description
Default
1
Select View Items
Select an item to be displayed. Select it with the small knob and press the
ENTER key. Every time the ENTER key is pressed, the check mark turns on
or off. Multiple checkboxes can be selected.
Current rms
Voltage rms
Effective value of input voltage
Voltage Peak +
Positive amplitude peak value of input voltage
Voltage Peak -
Negative amplitude peak value of input voltage
Current rms
Effective value of input current
Current Peak +
Positive amplitude peak value of input current
Current Peak -
Negative amplitude peak value of input current
Frequency
Input frequency measured with input voltage
Real Power
*1
(W)
Real power (W) of the EUT
Apparent Power
*1
(VA) Apparent power (VA) of the EUT
Reactive Power
*1
(var) Reactive power (var) of the EUT
Power Factor
*1
Power factor of the EUT
THC
Total harmonic current of input current, effective
value of all the harmonic current components from
the 2nd to 40th order
POHC
Odd-order harmonic current of input current,
effective value of odd-order harmonic current
components from the 21st to 39th order
THD
*2
Total harmonic distortion factor of input current,
effective value ratio of all the harmonic current
components from the 2nd to the 40th order against
the effective value of the fundamental components
THD-I
Total harmonic distortion factor of input current,
effective value ratio of rms current from the 2nd to
the 40th order against the effective value of the
fundamental components
PWHD
*2
Partial-weighted harmonic distortion factor of input
current, effective value ratio of all the harmonic current
components from the 14th to 40th order against the
effective value of the fundamental components
*1 In a three-phase three-wire connection, the
(sigma) values are the valid values. The values of each phase (L1 to L3)
carry no meaning, so they are displayed as "-----."
*2 In 61000-3-12 2011, THD becomes THC/Iref, and PWHD becomes PWHC/Iref. Iref is displayed as "-----," because it is
determined at the end of the test.
%
%
F1
Select View Items
Voltage rms
Voltage Peak +,
Voltage Peak -
Current rms
Frequency
Real Power (W)
Apparent Power (VA)
Reactive Power (var),
Current Peak +,
Current Peak -
POHC
Power Factor
F2
Scroll
(Input numeric value)
F5
L1
L2
L3
F6
Return
Check View Phase
THC,
PWHC/Iref
THD/Iref,
PWHD
THD,
1
3
2
4
61000-3-12 2011
61000-3-12 2004
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......