N9030B PXA Signal Analyzer Service Guide 85
Instrument Messages
Introduction
Event Message Format
The event messages are listed in numerical order according to their message
number. Advisory messages do not have numbers, and are listed in
alphabetical order.
An explanation is included with each error to further clarify its meaning. Some
errors are specified in industry standards and there are references to the IEEE
Standard 488.2-1992,
IEEE Standard Codes, Formats, Protocols and Common
Commands for Use with ANSI / IEEE Std 488.1-1987
. New York, NY, 1992.
Advisories
Advisory messages tell the front panel user some useful information. (For example, “File saved
successfully” or “Measuring the fundamental”.)
Advisory messages appear in the Status Panel at the bottom of the display. The message remains
until you press a key, or another message is displayed in its place.
Advisory messages are not logged in the error queues.
Grayout messages are a special type of Advisory, which appear when you attempt to access a
function that is not available. This could be a grayed out front panel key, or an inappropriate SCPI
command. There are two types of grayout messages: Benign and Forced.
1. Benign
: the requested function is not available because it does not make sense
with the current instrument settings. Changing it does not affect the current
measurement. (For example, setting the number of FFTs/Span when you are not
in the FFT mode.)
A benign grayout gives an Advisory type of message only when the front panel
key is pressed.
The requested function cannot be changed from the front panel, but it can be
changed remotely.
2. Forced
: the requested function is not available either because changing it
would cause an invalid measurement, or because of hardware limitations, or
because the selection conflicts with other settings. (For example, selecting the
electrical attenuator when the frequency span includes frequencies above 3.6
GHz.)
A forced grayout function cannot be changed either from the front panel or
remotely. It generates a special type of Advisory message. It also only appears
on the front panel when the key is pressed. Remotely, the message will appear
in the event queue as a warning “-221, Settings conflict; <conflict
description>”.
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...