230 N9030B PXA Signal Analyzer Service Guide
Analog/Digital IF Troubleshooting
A2 Analog IF Troubleshooting
A2 Analog IF Troubleshooting
There are three steps to verify the A2 Analog IF Assembly.
— Measure the input power and frequency accuracy of the 322.5 MHz signal
(from the A13 RF Front End assembly)
— Measure the output power and frequency accuracy of the 22.5 MHz signal
(to the A3 Digital IF Assembly)
— Measure the 300 MHz input LO signal (from the A16 Reference Assembly)
Verifying the 322.5 MHz Input Power
1.
Turn the instrument off.
2.
Remove instrument cover. Refer to
Chapter 16, “Assembly Replacement
3.
Turn on the PXA Signal Analyzer and wait for the instrument to complete
the boot up process.
4.
Press
System
,
Alignments
,
Auto Align
,
Off
.
5.
Press
Input/Output
,
RF calibrator
,
50 MHz
.
6.
Verify the 50 MHz signal is at
−
25 dBm on the analyzer display by pressing
FREQ
,
50 MHz
,
SPAN
,
1 MHz
,
Peak Search
on the PXA. The marker
readout should be 50 MHz at
−
25 dBm
±
3 dB. If this reference signal is
measuring incorrectly, see
Chapter 4, “RF Section Troubleshooting
7.
To continue verifying press
SPAN
,
Zero Span
. Verify the input attenuator
on the PXA is set to 10 dB. Look near the top of the display near the center
and verify that
Atten: 10 dB
is visible. If needed change the input
attenuator by pressing
AMPTD
,
Attenuation
,
10 dB
on the analyzer.
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...