N9030B PXA Signal Analyzer Service Guide 295
CPU/Disk Drive Troubleshooting
Disk Drive Recovery Process
Disk Drive Recovery Process
The Keysight Recovery System can be used to repair errors on the instrument's
C: drive partition or to restore the original factory configuration of the system
software. The Keysight Recovery System is stored in a separate hidden hard
disk drive partition.
Repairing errors on the hard disk drive may result in loss of data or files. If you
need more information about the Windows “chkdsk” error repair process, see
the chkdsk documentation in the Microsoft Windows Help and Support Center.
Restoring the original factory system software does not restore any of the
following items:
— Windows system configurations that were made after the instrument was
shipped from the factory. For example, Windows and Service Pack updates,
user accounts, and windows configuration settings. After an Keysight
Recovery, these configurations need to be redone.
— Additional software that was installed after the instrument was shipped
from the factory. After an Keysight Recovery, that software needs to be
re-installed.
— Any data or programs saved on the D: or E: drives.
— Any upgrades that were made to the Keysight measurement application
software.
Restoring the original factory system software will not synchronize the code in
the Field Programmable Gate Arrays (FPGAs) on the various hardware
assemblies. As a result, you may see an error dialog box appear during the final
boot-up at the end of the recovery process. This typically occurs when there
are significant differences between the software version installed prior to
performing the recovery and the version restored by the recovery. In these
situations, upgrade the software to the latest version.
Calibration data is not lost during this process because it resides on the E drive
partition.
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...