N9030B PXA Signal Analyzer Service Guide 139
Instrument Messages
Condition Messages
309
4
User Cal; Freq outside cal
range
E
The existing user cal has been invalidated
because the current measurement
frequencies lie partially or wholly outside
the range of frequencies used for
user-cal. (UNCAL)
309
4
User Cal; Cal will be
interpolated
W
The measurement frequency range has
been changed such that it is a subset of
the calibrated range. (~CAL)
309
4
User Cal; Adjusted for new
RBW
W
The measurement RBW has been
changed since the last calibration (~CAL)
311
5
Calibration
W
311
5
Calibration; ENR table
extrapolated
W
One or more calibration or measurement
frequency points exceed the currently
loaded Cal or Meas ENR Table frequency
ranges. The corresponding ENR table’s
lowest frequency ENR value will be
re-used for frequencies less than the
table range, and the highest frequency
ENR value will be re-used for frequencies
greater than the table range.
(~ENR)
311
5
Calibration; No ENR data
present
W
No ENR Data (ENR)
313
6
Source Uncal
W
313
6
Source Uncal;adj Start Freq
or RBW
W
While using a Tracking Source, you must
make sure the Start Frequency is high
enough to avoid capturing LO
feedthrough in the trace. This depends
on both Start Freq and RBW. If you get
this message, increase your Start Freq or
narrow your RBW.
315
7
unused W
317
8
unused W
319
9
unused W
321
10
unused W
323
11
unused W
325
12
unused W
327
13
unused W
Err#
Status
Register
Bit
Message
Error or
Warning
Verbose/Correction Information
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...