42 N9030B PXA Signal Analyzer Service Guide
Overview
Service Equipment You Will Need
Step Attenuator Loss Characterization
The step attenuator combination should have each attenuator setting
characterized by a metrology lab at 50 MHz. This characterization can be
ordered through Keysight Technologies as an Option H50 Calibration.
The following tables show which sections of the 10 dB and 1 dB step
attenuators are utilized for each attenuator setting. The tables also list the
Recommended Uncertainty for each attenuator setting. A larger number will
result in larger overall test uncertainties which could affect the test pass/fail
rate.
d. Wenzel Associates Inc
2215 Kramer Ln
Austin, Texas 78758
(512) 835-2038
FAX (512) 719-4086
e. The 8482A power sensor uses cal factors to compensate the power sensor for frequency
response errors. Cal factors are stated in percentages. The 8482A factory cal factor uncertainty
ranges from 2.2% to 3.1%. The cal factor uncertainty can be reduced to < 2.0% by using metrol-
ogy grade calibration techniques. The power sensor cal factor uncertainty becomes one compo-
nent of the Verification Test uncertainty analysis. Lower cal factor uncertainties will translate to
wider test margins.
f. When ordering a new STD or CFT Power Sensor from Keysight to be used with the N7800A series
applications, order with Option 1A7. The 1A7 option provides an ISO17025 calibration and
includes calibration data. When re-calibrating your STD or CFT Power Sensor for use with the
N7800A series applications, we recommend the Keysight ISO17025 calibration.
g. When ordering a new STD or CFT Power Sensor from Keysight to be used with the N7800A series
applications, order with Option 1A7. The 1A7 option provides an ISO17025 calibration and
includes calibration data. When re-calibrating your STD or CFT Power Sensor for use with the
N7800A series applications, we recommend the Keysight ISO17025 calibration.
h. The step attenuators should be permanently joined via the 11716A Interconnect Kit as shown in
the diagram.
i. 8491A Option H33 is a fixed attenuator which has been characterized to have a VSWR
≤
1.05:1 at
50 MHz. A VSWR of 1.05:1 is critical to test Input Attenuator Switching Uncertainty, Display Scale
Fidelity, and Absolute Amplitude Accuracy performance tests. Any 8491A/B attenuator can be
mapped into this device if the VSWR at 50 MHz has been characterized to be
≤
1.05:1. When
mapping the attenuator, to indicate that is meets the required specification, the Option H33
checkbox will need to be checked in the configure test station Administration screen.
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...