334 N9030B PXA Signal Analyzer Service Guide
Hardware Options
A25 Wideband Analog IF Troubleshooting - Option B2X and B5X
A25 Wideband Analog IF Assembly Troubleshooting - Option B2X and
B5X
Verify the J17 IF Input level
1.
Turn the instrument off.
2.
Remove the instrument enclosure and top brace covers. Refer to
16, “Assembly Replacement Procedures”
3.
Turn on the signal analyzer and wait for the instrument to complete the
boot up process.
4.
Press
System
,
Alignments
,
Auto Align
,
Off
.
5.
Press
Mode/Meas
,
IQ Analyzer (Basic)
,
Complex Spectrum
,
OK
6.
Press
Input/Output
,
RF Calibrator
,
50 MHz
7.
Press
Frequency
hard key and set the center frequency to 50 MHz.
8.
Press
Meas Setup
,
IF Path
. Tap
IF Path
and choose
255 MHz
or
510 MHz
from the drop down menu. If Option B5X, 510 MHz, is installed, there is no
255 IF Path selection. Choose
510 MHz
in this case.
9.
Verify the 50 MHz signal is at
−
25 dBm on the analyzer display by pressing
Peak Search
on the analyzer. The marker readout should be 50 MHz at
−
25 dBm
±
. If this reference signal is measuring
incorrectly, see
Chapter 4, “RF Section Troubleshooting (RF/Microwave
in this service guide.
Figure 12-3
50 MHz Signal
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...