292 N9030B PXA Signal Analyzer Service Guide
CPU/Disk Drive Troubleshooting
A5 Disk Drive
Overview
The A5 Disk Drive assembly has been divided up into four different partitions.
They are:
In addition, a hidden recovery partition is located on the drive. This partition
contains an image of the C: drive as it was when the instrument was shipped
from the factory. To restore the C: drive using the image stored in this recovery
partition see the section titled
“Disk Drive Recovery Process” on page 295
Table 9-1
C.
This partition contains the operating system and software installed by Keysight. This is
an open system which means you can install additional software, which should be
installed on the C: drive. However, only a limited set of software applications are tested
for use with the instrument software. The installation and/or use of other software is not
warranted, and could interfere with the operation of the instrument software. If the
Keysight Recovery process is ever run, the original version of the C: drive, as shipped
from the factory, will be restored. The user will need to reload any other software that
was previously installed into the instrument.
Do not save any user data to the C: drive partition, as any data saved in this partition will
be lost if the Keysight Recovery process is run.
D.
This partition is reserved for data storage. The User Accounts that are preconfigured by
Keysight and their My Documents folder are mapped to the D: drive. This is for the
convenience of backing-up the user data. You should always back-up the data on the D:
drive. This allows you to restore the data if the A5 Disk Drive assembly ever needs to be
replaced.
Data saved in this partition will not be lost if the Keysight Recovery process is run.
E.
This partition is reserved for Keysight's use. The primary use of the E: drive is for storing
of the instrument Calibration and Alignment data. Do not change or overwrite the files
on this drive. This could cause your instrument to not meet specifications, or even to
stop functioning correctly. It is also recommended that you back up the contents of this
drive. This allows you to restore the data if the A5 Disk Drive assembly ever needs to be
replaced, which could otherwise require that all instrument adjustments be performed.
While data saved in this partition will not be lost if the Keysight Recovery process is run,
do not use this drive for data storage.
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...