154 N9030B PXA Signal Analyzer Service Guide
RF Section Troubleshooting (RF/Microwave Analyzers)
Troubleshooting
Troubleshooting
Quick Check to Verify the Low Band Signal Path
The analyzer has an internal 50 MHz amplitude reference signal that is used to
verify the low band path. This 50 MHz calibrator is used when the analyzer
performs many of the internal alignment routines. Therefore, the functionality
of the 50 MHz calibrator and low band path are critical to passing Align All
Now. Refer to the RF Lowband Path Block Diagram (Options 503, 508, 513,
and 526) in
for details. Reference the instrument settings box on
the block diagram.
Equipment needed:
Spectrum Analyzer with frequency range to 8 GHz
Signal source with output power 0 dBm and frequency range to at least
3.6 GHz
Power splitter, 11667B
2 each, quality SMA or 3.5 mm cables
90
°
SMA (m) to SMA (f) adapter
MMCX (f) to SMA (f) cable
SMA (f) to SMA (f) adapter
SMA (m) to SMA (m) adapter
Turn the instrument on and allow it to complete its full boot up process to
Spectrum Analyzer mode. Use the internal 50 MHz,
−
25 dBm calibrator signal
as a reference for troubleshooting by pressing
Mode Preset
,
Input/Output
,
RF
Calibrator
,
50 MHz
on the analyzer. Now press
FREQ (Channel)
,
50 MHz
,
SPAN (X Scale)
,
1 MHz
,
Peak Search
. The 50 MHz reference signal should
measure 50 MHz @
−
25 dBm
±
2 dB on the analyzer display. If the power level
is within tolerance, the low band path is functioning correctly at 50 MHz. If the
power level is incorrect, select the 4.8 GHz RF Calibrator and set the analyzer
center frequency to 4.8 GHz. Since several of the same signal path
components are shared between low band and high band, determining if the
problem appears in both paths is helpful.
To verify the high band path go to
“Quick Check to Verify High Band RF Path”
If the power level is not within tolerance, press
SPAN (X Scale)
,
Zero Span
,
AMPTD (Y Scale)
,
Attenuation
,
10 dB
.
To perform the following checks, it will be necessary to remove the outer
cover and the top brace. See
for removal procedures.
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...