N9030B PXA Signal Analyzer Service Guide 569
Post-Repair Procedures
Post-Repair Procedures
FPGA Synchronization
There is FPGA (Field-Programmable Gate Array) program code on many
different assemblies inside of the instrument, and all of these must be of a
compatible version with the instrument software revision. In order to ensure
that this requirement is met when either an assembly is replaced that contains
FPGA code on it, or if the A5 Disk Drive is replaced with a drive that has a
different version of software than the previous drive, you will need to take
certain steps to ensure that the instrument operates properly.
Updating the Instrument FPGA Code
All instruments have a utility included in them that will update the FPGA code
on all of the different assemblies in the instrument that have it. When this
utility is run it will detect the assemblies that need to be updated and will
update them to the correct revision.
1.
Close the instrument application software by pressing
File
,
Exit
,
Enter
using the front panel keys.
2.
Connect a USB mouse to one of the instrument front panel USB ports.
3.
Using the mouse, click
Start
,
Log Off
,
Log Off
.
4.
Re-login as “administrator”. The default password is “Keysight4u!”. Click
OK
.
5.
Using the mouse double-click on the Windows Explorer link on the
desktop and navigate to the following folder:
C:\Program Files\Agilent\SignalAnalysis\Physics
6.
In this folder find and execute the file named:
FPGA_Prog.bat
7.
The FPGA Programming Utility will start and a window as shown in
will appear.
Even if an instrument appears to work properly when an assembly with
FPGA code is replaced, that does not mean that the FPGA code is
completely compatible with the software version installed in the
instrument. There could be incompatibilities that will only be seen under
certain measurement conditions.
Once you start the FPGA programming process you MUST NOT interrupt
the process for any reason. This would include turning the instrument off
or unplugging the power cord to the instrument. Doing so will result in an
inoperative instrument, requiring the affected assembly to be replaced.
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...