326 N9030B PXA Signal Analyzer Service Guide
Hardware Options
A25 Wideband Analog IF Troubleshooting - Option B1X and B85
Inputs to the A25 Assembly
A25J101 300 MHz IF signal (from A15J901 Front End Controller)
A25J301 2400 MHz reference from A16 Reference Assembly via W34 cable
A25J805 Step Calibrator from A3J17 40 MHz Digital IF via W45 cable
Outputs from the A25 Assembly
A25J302 2400 MHz Reference Out
A25J102 250 MHz IF Out to A3J15 of the 40 MHz Digital IF
A25J806 50 MHz Comb Out to A16 Reference assembly A16J726 via W46
cable
A25J2 ADC Data to 160 MHz Digital IF A26J504 via W44 cable
A25 Wideband Analog IF Assembly Theory of Operation - Option B1X
and B85
The A25 Wideband Analog IF assembly primary role is to properly filter and
apply attenuation or gain to optimize the level going to the ADC and provide
sufficient pre-filtering to reject images that would otherwise interfere with the
signal of interest. Precision alignments are used to calibrate the filters,
attenuators and gain of this assembly. It is not uncommon that the reported
failure will come from failures reported when an alignment fails for this
assembly. The alignment history file is a good indicator of a failure for this
assembly. This file is a text file that can be edited with Notepad. The file is
located in the following location:
E:\AlignDataStore\AlignmentHistory.txt
This file will list all alignments that have passed or failed. Alignments reporting
a failure of the WBAIF assembly can in many cases be attributed to failure of
this assembly. If however these errors follow errors in the RF section of the
instrument, it is possible that additional WBAIF error will appear due to
missing RF Calibration signals and not related to the A25 Wideband IF
assembly.
Manually troubleshoot the assembly is basically verifying the 300 MHz IF signal
is present and at the proper level. In addition you will want to verify input and
output levels of the various calibration and reference signals.
Additionally, the A25 Analog IF has a built in RF Burst Trigger.
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...