196 N9030B PXA Signal Analyzer Service Guide
RF Section Troubleshooting (Millimeter-Wave Analyzers)
Troubleshooting
the analyzer LO phase lock loop to unlock, which would affect the LO
frequency measurement. The LO signal should measure 5122.5 MHz at +10
dBm
±
3 dB.
If the signal level is incorrect, measure the signal coming from the A20 YTO at
A13A1J4. In order to measure this signal and avoid an LO Unlock, you will need
to install a power splitter at A13A1J4. Connect one arm of the splitter to
A13A1J4 using a SMA (m) to SMA (m) adapter. Connect the input of the
splitter to W20 using the SMA 90 degree adapter and short SMA cable.
Connect the measuring spectrum analyzer to the other arm of the splitter. Add
the loss through the splitter arm which is 6 dB for the Keysight 11667B or
11667C splitter. The signal should be 5122.5 MHz at +10 dBm
±
3 dB.
A13 Front End Assembly Verification
Many portions of the Front End Assembly have been verified earlier in the High
Band troubleshooting process outlined above.
The input signal level was measured on W15 as part of the A13 Front End Input
Verification.
The output signal level was measured at A13A1J7 during the quick check to
verify the High Band signal path.
The LO input at W20, and the LO outputs were tested when performing the
First LO Verification.
High Band Preamp (Option P08, P13, P26, P43, P44, P50)
If a preamp option listed above is installed, the High Band preamp can be
verified as follows:
View the 4.8 GHz calibrator signal on screen. Press
AMPTD
,
More
,
Internal
Preamp
,
Full Range
to activate the preamp.The expected operation is the
signal level will not change more than +/- 0.5 dB and the noise floor will
increase due to the fact the input attenuation was automatically increased.
If the amplitude error when switching the preamp on and off is greater than 0.5
dB, assure that any other internal alignment failures are resolved since the
System Gain alignment must run before optimum performance is possible. The
amplitude error could be caused by poor frequency response, and performing
the high band frequency response adjustment (using the calibration software)
will adjust the preamp on, high band path. Assure you have evaluated all
causes including A15 Front End Controller switching problems, before you
replace the Front End assembly.
High Band Mixing Equations
For input signal frequencies from 3.6 GHz to 8.4 GHz:
The A13 Front End assembly is shown as A13A1 and A13A2 for
troubleshooting clarification. However, A13A1 and A13A2 are not
separately replaceable. The entire A13 assembly must be replaced.
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...