280 N9030B PXA Signal Analyzer Service Guide
CPU/Disk Drive Troubleshooting
A4 Processor Board Assembly Troubleshooting
A4 Processor Board Assembly Troubleshooting
The A4 Processor board assembly is serviced as an assembly only; no
component level repair is supported.
Boot-Up or Initialization Problems
Typical failures of the A4 Processor board assembly will cause the instrument
to not boot-up or initialize properly. These types of failures can also be caused
by a variety of other assemblies as well. In order to determine whether a
problem such as this is being caused by a defective A4 Processor board
assembly, first see the PXA Instrument Boot Up Process on page 53 to
eliminate other possibilities.
Remote Monitor Troubleshooting
When an external monitor is connected to the analyzer, the remote monitor
should automatically begin working.
If the remote monitor does not function, perform the following:
1.
Turn off the analyzer.
2.
Assure the external monitor has power and that the VGA cable is
connected properly to both the monitor and the analyzer.
3.
Turn on the analyzer and wait for it to fully boot-up.
4.
Connect a mouse to the analyzer, and either click on the desktop, or on
the blue display icon on the task bar. Choose
Graphics Properties
.
5.
Compare the settings on your analyzer to the settings shown below:
Figure 9-1
Monitor Set-Up
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...