Instrument Messages
Event Messages
–221
Settings conflict; Gate is
not available when Marker
Count on
The gate function cannot be used while you have marker
count turned on.
–221
Settings conflict; Gate
Length is not settable in
FFT sweeps
The sweep time for FFT sweeps is set by the calculations. So
sweep time settings cannot be adjusted.
–221
Settings conflict; Gate
Length is not settable in
FFT sweeps
The sweep time for FFT sweeps is set by the calculations. So
sweep time settings cannot be adjusted.
–221
Settings conflict; Gate
Method is not compatible
with current Sweep Type
setting
If the Gate is On and you have the FFT Sweep Type manually
selected, then the Gate Method cannot be selected.
–221
Settings conflict; Gate not
available with external
Tracking Source
The Gate functions are unavailable when Source Mode is
Tracking with an external source. This is because the Gate
circuitry is used to sync the external source.
–221
Settings conflict; Gate not
available with Tracking
Generator
If the Source Type is Tracking Generator, the Gate circuitry is
used for TG sync and is not available for gating
–221
Settings conflict; Gated FFT
is not available while Sweep
Type is set to Swept
The gated FFT function is not available if you have selected
the swept type of sweep. You must be in the FFT sweep type.
–221
Settings conflict; Gated LO
is not available while Sweep
Type is set to FFT
The FFT sweep type moves the LO frequency in steps. So the
gated LO function is not available if you have selected FFT
sweep.
–221
Settings conflict; Gated
Video is not available while
Sweep Type is set to FFT
The gated video function is not available if you have selected
the FFT sweep type.
–221
Settings conflict; Incorrect
RBW for demod. Change RBW
–221
Settings conflict; Ind I/Q
is not available for this
measurement
The Independent I and Q setting is not available for the
current measurement. Only some measurements (initially,
only VXA) support this setting.
–221
Settings conflict; Invalid
trace number
The subopcode used to specify the trace number is invalid
for this measurement or query
–221
Settings conflict; Knob is
not available to modify this
function
You should select a specific value for this function. So
scrolling through values with the knob is not allowed.
–221
Settings conflict; LO Phase
Noise Adj not available
For instruments without the Dual-Loop LO, this feature is not
available
Err#
Message
Verbose/Correction Information
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...