N9030B PXA Signal Analyzer Service Guide 157
RF Section Troubleshooting (RF/Microwave Analyzers)
Troubleshooting
Troubleshooting a Low Band Problem
Refer to the RF Lowband Path Block Diagram in
and follow the
instructions in the settings box. To enable the internal 50 MHz,
−
25 dBm
calibrator signal press
Input/Output
,
RF Calibrator
,
50 MHz
.
The Low Band signal path (sometimes referred to as Band 0) is used for all
signals less than 3.6 GHz when the analyzer is used in normal operation. If the
stop frequency is set to 3.6 GHz, the analyzer displays only the Low Band
signal path. If the analyzer stop frequency is set above 3.6 GHz and the start
frequency is set below 3.6 GHz, the analyzer will sweep up to 3.6 GHz in low
band and then automatically switch to the high band path above 3.6 GHz.
A11 Low Band Switch Verification
Disconnect W8 from A13A1J2. See
for location of connector. Attach
a right angle sma adapter to W8 and measure the output of the Low Band
Switch with a spectrum analyzer. Expected signal is 50 MHz at
−
35.3 dBm
since the Low Band Switch and Option LNP SW3 have a few tenths of a dB
loss.
If the signal level is incorrect, remove A10 Input Attenuator B output cable and
measure the output power. Expected signal is 50 MHz at
−
35 dBm. See
for location of attenuator. If the attenuator output signal is correct but the
Low Band output at W8 is incorrect, suspect the A11 Low Band Switch, the
interconnect cables, Switch 3 if Option LNP is present, or the control signals.
The control signals are explained in
.
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...