216 N9030B PXA Signal Analyzer Service Guide
Front End Control Troubleshooting
A15 Front End Control Assembly Troubleshooting
Verifying SW3 (Option LNP only)
Since there is a ribbon cable from the A15J802 to SW3, it is not feasible to use
a voltmeter or oscilloscope to measure the control voltages. For testing SW3,
the “click” test can be used.
Press
Mode Preset
on the analyzer. Press
FREQ
,
5 GHz
,
SPAN
,
1 MHz
. Verify
the optional preamplifier is off by pressing
AMPTD
,
More 1 of 2
. The Internal
Preamp softkey must be Off in order to test SW3. Press
AMPTD
,
More 1 of 2
,
uW Path Ctrl
,
Low Noise Path Enable
on the analyzer. You should hear a
distinct “click” noise from SW3 when the low noise path is selected. Now press
uW Path Ctrl
and select
Standard Path
. SW3 will “click” again and return to
the standard RF path.
Another way of testing this switch is using an RF signal. This is described in the
RF Section Troubleshooting chapter, under the high band troubleshooting
section.
Verifying SW1 and SW2 (Option MPB only)
SW1 and SW2 can be verified three different ways:
— “Click” Test
— Oscilloscope
— R.F. Signal Tracing
“Click” Test
This test changes between two different path modes. When this is done a
distinct “click” noise from SW1 and SW2 can be heard. Since both switches
switch at the same time, it is difficult to decipher if one is switching and the
other is not.
Press
Mode
,
Preset
on the analyzer. Press
FREQ
,
5 GHz
,
SPAN
,
1 MHz
,
AMPTD
,
More 1 of 2
,
uW Path Ctrl
. Standard Path is the default. Select uW
Preselector Bypass and you should hear a distinct “click” noise if the control
logic from the A15 Front End Control board is getting to SW1 and SW2.
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...