N9030B PXA Signal Analyzer Service Guide 593
Functional Tests
Frequency Response (Flatness), Preamp On
Frequency Response (Flatness), Preamp On
Test Limits
The frequency response test, with preamplifier on, measures the signal
analyzer’s amplitude error as a function of the tuned frequency. Measurements
are made ranging from 50 MHz to the maximum frequency range of the
preamp. The signal source amplitude is measured with a power meter to
eliminate error due to source flatness. The measured value is normalized to 50
MHz.
Frequency Range
Limit Relative to 50 MHz
100 kHz to 3.6 GHz
±2.0 dB
> 3.6 GHz to 8.4 GHz
±3.0 dB
> 8.4 GHz to 13.6 GHz
±3.5 dB
> 13.6 GHz to 26.5 GHz
±4.0 dB
> 26.5 GHz to 34.5 GHz
±4.0 dB
> 34.5 GHz to 50 GHz
±5.0 dB
Item
Critical Specifications
(for this test)
Recommended
Keysight Model
Adapter
Type N (m) to 3.5 mm (m)
Frequency: 10 MHz to 18 GHz
VSWR: < 1.1:1
1250-1743
Adapter
3.5 mm (f) to 3.5 mm (f)
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.1:1
83059B
Cable, 3.5 mm, 1 meter
(Opt P03, P08, P13, P26)
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.4:1
11500E
Cable, 2.4 mm, 1 meter
(Opt P43, P44, P50)
Frequency: 10 MHz to 50 GHz
VSWR: < 1.55:1
8120-6164
Cables, BNC
120 cm
Frequency: 10 MHz
10503A
Power Meter
Compatible with power sensor
E4418B
Attenuator, 30 dB Fixed
Frequency: 50 MHz
Accuracy: ±0.05 dB
11708A
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...