N9030B PXA Signal Analyzer Service Guide 113
Instrument Messages
Event Messages
–222
Data out of range; Two
entries already exist at
this x-axis value.
When entering values for limit lines, you cannot have more than
two y-axis (amplitude) values entered for a specific x-axis
(frequency) value.
–223
Too much data
A data element (of block, expression, array type, or string type)
had more data then allowed by the command, or by the available
memory.
–223
Too much data; 200 spurs
found. Additional spurs
ignored.
There are too many spurs for the table (the limit is 200), and any
additional spurs that are found will be ignored.
–224
Illegal parameter value
An exact data value (from a list of the allowed values) was required
- but not found. See the feature description for information about
the expected parameter values.
–224
Illegal parameter value;
<Value> invalid.
Fractional values are not
allowed.
The seconds parameter of an LXI time may not contain a fractional
portion. For example 123456789.0 is valid while 123456789.1 is
not.
–224
Illegal parameter value;
<value> out of range.
The value does not fall in the valid range
–224
Illegal parameter value;
Exceeding the max list
length
The list parameters have a maximum allowed length. You are
trying to set a length longer than the maximum.
–224
Illegal parameter value;
Gated FFT is not available
while Sweep Type is set to
Swept
The gated FFT function is not available if you have selected the
swept type of sweep. You must be in the FFT sweep type.
–224
Illegal parameter value;
Gated LO is not available
while Sweep Type is set to
FFT
The FFT sweep type moves the LO frequency in steps. So the
gated LO function is not available if you have selected FFT sweep.
–224
Illegal parameter value;
Gated Video is not
available while Sweep Type
is set to FFT
The gated video function is not available if you have selected the
FFT sweep type.
–224
Illegal parameter value;
Illegal identifier
<identifier>. This value
may already be in use.
The value for the LXI LAN identifier parameter must be unique (i.e.
LAN0 and LAN7 must have different identifier strings).
–224
Illegal parameter value;
Index out of range
When querying the LXI Event Log or the Servo Log, an index may
be used to look at a specific entry. This error occurs if the index
provided does not point to a valid entry.
Err#
Message
Verbose/Correction Information
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...