N9030B PXA Signal Analyzer Service Guide 115
Instrument Messages
Event Messages
–230
Data corrupt or stale;
Unable to load state from
file
There is something wrong with the state data in the desired file.
Maybe the file is corrupt, or it is from an instrument/version that is
not recognized by the current instrument.
–231
Data questionable
Indicates that the measurement accuracy is suspect
–232
Invalid format
A data element was found but it could not be used because the
data format or the data structure was not correct.
–232
Invalid format; Map
information not loaded
Instrument failed to load the burst mapping information from the
selected file.
–233
Invalid version
A legal data element was found but could not be used because
the version of the data is incorrect. For example, state data
changes as new instrument features are added, so old state files
may not work in an instrument with a newer version of software.
–240
Hardware error
A legal program command or query could not be executed
because of a hardware error. The exact problem cannot be
specifically identified.
–240
Hardware error; See
details in Windows Event
Log under SA
The internal data acquisition system detected a problem at
startup and logged the details in the Windows Event Log.
–241
Hardware missing
The operation could not be performed because of missing
hardware; perhaps the optional hardware is not installed.
–241
Hardware missing; Internal
preamp not available at
all frequency points
The Internal Preamp is currently turned on, but the measurement
is being performed completely or partially outside the range of the
preamp. It is recommended that the user turns preamp off to
ensure consistent results across the entire measurement.
–241
Hardware missing; not
available for this model
number
The hardware required is not part of this model
–241
Hardware missing; Option
not installed
The optional hardware is not installed.
–250
Mass storage error;
A problem was found with the mass storage device (memory, disk
drive, etc.). The exact problem cannot be specifically identified.
–250
Mass storage error; Access
denied
Access is denied.
–250
Mass storage error; Bad
path name
The specified path is invalid.
–250
Mass storage error; Can
only import single trace
.csv files
Trace files containing multiple traces can not be imported.
However, if you need to recall multiple traces you can use the
Save and Recall functions rather than the Import and Export
functions.
Err#
Message
Verbose/Correction Information
Содержание N9030B
Страница 4: ...4 ...
Страница 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Страница 172: ...172 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting RF Microwave Analyzers Troubleshooting ...
Страница 198: ...198 N9030B PXA Signal Analyzer Service Guide RF Section Troubleshooting Millimeter Wave Analyzers Troubleshooting ...
Страница 248: ...248 N9030B PXA Signal Analyzer Service Guide Analog Digital IF Troubleshooting A3 Digital IF Troubleshooting ...
Страница 298: ...298 N9030B PXA Signal Analyzer Service Guide CPU Disk Drive Troubleshooting Disk Drive Recovery Process ...
Страница 320: ...320 N9030B PXA Signal Analyzer Service Guide Front Panel Motherboard Troubleshooting A1 Front Panel Assembly ...
Страница 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Страница 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 356: ...356 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Analog IF Block Diagram ...
Страница 357: ...N9030B PXA Signal Analyzer Service Guide 357 Block Diagrams Block Diagrams Digital IF Block Diagram ...
Страница 358: ...358 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams PXA Computer Block Diagram ...
Страница 402: ...402 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 6 RF Area Options 503 508 513 526 ...
Страница 404: ...404 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 7 RF Area Options 544 550 ...
Страница 418: ...418 N9030B PXA Signal Analyzer Service Guide Replaceable Parts Hardware Figure 15 17 AIF Cables Figure 15 18 DIF Cables ...
Страница 427: ...N9030B PXA Signal Analyzer Service Guide 427 Replaceable Parts Hardware Figure 15 26 Front Frame Exploded View ...
Страница 522: ...522 N9030B PXA Signal Analyzer Service Guide Assembly Replacement Procedures AIF DIF Assembly Figure 16 85 AIF Cables ...
Страница 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Страница 608: ...608 N9030B PXA Signal Analyzer Service Guide Functional Tests BBIQ Input Frequency Response Option BBA only ...
Страница 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...